Table 1

Acute studies

Pig numberComplications
Pig 1Kinking of the loops in the LV. Haematoma in the annulus of the mitral valve. Device position too low, ie, located in the IVC
Pig 2Kinking of the loop. Device caught by Eustachian valve
Pig 3Pericardial effusion during puncture of the atrial septum. Kinking of the loop. Damage to aortic valve
Pig 4Died from ventricular fibrillation
Pig 5Separation of the two umbrellas a few seconds after implantation. Retraction of the right atrial umbrella. Embolisation of the left atrial umbrella to the aortic valve. Corresponding to a weld on the metallic skeleton, a fracture of one arm in this umbrella had occurred. This was the only pig in which we intended to implant a 45 mm oversized device
Pig 6Pre-existing ASD. Too big for closure with a 35 mm device. Successful retrieval of the device was performed
Pig 7Pre-existing ASD. Too big for closure with a 35 mm device. Successful retrieval of the device was performed
Pig 8Laceration of the inguinal ligament and the femoral artery
Pig 9Laceration of the inguinal ligament and the femoral artery
  • ASD, Atrial septal defect; IVC, Inferior vena cava; LV, Left ventricle