Pig number | Complications |
---|---|
Pig 1 | Kinking of the loops in the LV. Haematoma in the annulus of the mitral valve. Device position too low, ie, located in the IVC |
Pig 2 | Kinking of the loop. Device caught by Eustachian valve |
Pig 3 | Pericardial effusion during puncture of the atrial septum. Kinking of the loop. Damage to aortic valve |
Pig 4 | Died from ventricular fibrillation |
Pig 5 | Separation of the two umbrellas a few seconds after implantation. Retraction of the right atrial umbrella. Embolisation of the left atrial umbrella to the aortic valve. Corresponding to a weld on the metallic skeleton, a fracture of one arm in this umbrella had occurred. This was the only pig in which we intended to implant a 45 mm oversized device |
Pig 6 | Pre-existing ASD. Too big for closure with a 35 mm device. Successful retrieval of the device was performed |
Pig 7 | Pre-existing ASD. Too big for closure with a 35 mm device. Successful retrieval of the device was performed |
Pig 8 | Laceration of the inguinal ligament and the femoral artery |
Pig 9 | Laceration of the inguinal ligament and the femoral artery |
ASD, Atrial septal defect; IVC, Inferior vena cava; LV, Left ventricle