Complications of pacemaker-defibrillator devices: Diagnosis and management

https://doi.org/10.1016/0002-8703(94)90090-6Get rights and content

Abstract

Treatment of resuscitated patients with implantabie cardioverter defibrillators has become increasingly more common as a method for the prevention of sudden cardiac death. Major complications such as perioperative death (incidence 2% to 8%), infection (2% to 11%), and lead-related problems (3% to 27%) have been described in previous trials. In our experience with 140 patients, problems were related to leads (n = 11), the device (n = 2), pacing (n = 1), sensing (n = 13), and defibrillation function (n = 5). Additional problems that occurred during the perioperative period included infection (n = 11), hematoma, and seroma (n = 2). Thrombus formation along endocardial leads was observed in 13 of 62 (21%) patients. Different arrhythmias (n = 10), such as sinus tachycardia, atrial fibrillation, and nonsustained, slow or incessant ventricular tachycardia with shock delivery, were also detected. Surgical management (predominantly for the major problems) was used in 31 (48%) patients, drug treatment in 25 (39%), and reprogramming of the device in 24 (38%) patients. All of these problems can results in an increase in mortality rates. This article provides an overview of the complications of cardioverter defibrillator treatment and is based on both published data and our series.

References (38)

  • H Klein et al.

    Programmable implantable defibrillator—progress in defibrillator therapy or technical luxury?

  • FE Marchlinski et al.

    The automatic implantable cardioverter-defibrillator: efficacy, complications, and device failures

    Ann Intern Med

    (1986)
  • M Meesmann

    Factors associated with implantation-related complications

    PACE

    (1992)
  • S Saksena et al.

    The PCD investigators and participating institutions. Complications of third-generation defibrillators using endocardial or epicardial leads: a multicenter study [Abstract 742-6]

    J Am Coll Cardiol

    (1993)
  • JM Kleman et al.

    An intention to treat analysis of transvenous and epicardial ICD system implantation [Abstract 100]

    PACE

    (1993)
  • PFA Bakker et al.

    Infections involving implanted cardioverter defibrillator devices

    PACE

    (1992)
  • M Fromer et al.

    Efficacy of automatic multimodal device therapy for ventricular tachyarrhythmias as delivered by a new implantable pacing cardioverter-defibrillator: results of a European multicenter study of 102 implants

    Circulation

    (1992)
  • J Borbola et al.

    The automatic implantable cardioverter-defibrillator: clinical experience, complications, and follow-up in 25 patients

    Arch Intern Med

    (1988)
  • RL Taylor et al.

    Infection of an implantable cardioverter defibrillator: management without removal of the device in selected cases

    PACE

    (1990)
  • Cited by (68)

    • Management of cardiac device-related infections: A review of protocol-driven care

      2013, International Journal of Cardiology
      Citation Excerpt :

      The mean cost for a hospitalization attributed to a CIED infection ranges from $31,149 to $55,003; moreover, CIED infections are associated with an 8.4 to 11.6 fold increase in mortality when compared to hospitalizations attributed to noninfectious, cardiac device complications [5]. Device infections are one of the leading indications for extraction [6]. The term device infection can be subdivided into several categories, depending on what anatomical structures or device components are affected.

    • Intracardiac thrombus: Diagnosis, complications and management

      2013, American Journal of the Medical Sciences
    • Nosocomial Escherichia coli pacemaker pocket infection

      2011, American Journal of Infection Control
    • Complications of ICD Generator Change and Implantations

      2011, Cardiac Electrophysiology Clinics
      Citation Excerpt :

      ICD infections may be classified as localized pocket infections, erosion, or lead-related endocarditis (Fig. 7, video 3, lead vegetation, at www.cardiacEP.theclinics.com). Depending on whether both pacemakers and ICDs are included, the length of follow-up, the location of the pulse generator, transvenous versus epicardial lead positions, and concomitant cardiac surgery, the CIED infection rate ranges between 0.03% and 7.9%.23,24,28,34,42–53 Although the number of initial CIED implantations has risen, it is exceeded by the rate of rise of CIED infections as a percentage of new CIED implantations.73,74

    View all citing articles on Scopus