Estimation of effective dose to the patient during medical X-ray examinations from measurements of the dose-area product

Published under licence by IOP Publishing Ltd
, , Citation J C Le Heron 1992 Phys. Med. Biol. 37 2117 DOI 10.1088/0031-9155/37/11/008

0031-9155/37/11/2117

Abstract

Relationships between effective dose and each of entrance surface dose, entrance air kerma and dose-area product were studied for common radiographic projections using the organ dose data from Monte Carlo modelling. Dose-area product proved to be the best quantity for estimating effective dose and these data are presented. For typical peak generating potentials, estimates of effective dose can be made from dose-area product using a small number of conversion coefficients. For cervical spine, chest, kidneys, abdomen, lumbar spine and pelvis AP projections, the conversion coefficient 0.21 mSv per Gy cm2 will estimate effective dose within 30% of the value obtained using specific organ doses. Conversion coefficients for lateral and PA projections of the trunk are approximately half this value. These coefficients allow the direct estimation of effective dose with sufficient accuracy for most purposes.