Abstract
Magnetic resonance imaging (MRI) is a versatile noninvasive tool for achieving full-field quantitative visualization of biomedical fluid flows. In this study, two MRI velocimetry techniques (spin tagging and phase contrast) are used to obtain velocity measurements in a Poiseuille flow for Reynolds numbers below 1000. Spin-tagging MRI velocimetry supplies the displacement of tagged grids of nuclear spins from which the velocity field can be inferred, while phase contrast MRI velocimetry directly provides velocity data for every pixel in the field of view. Although the phase contrast method is more accurate for this flow, this technique is more sensitive to errors from magnetic susceptibility gradients, higher order motions, and has limited dynamic range. Spin-tagging MRI velocimetry is a viable alternative if automatic methods for extracting velocity fields from the tags can be found. Optical flow, a technique originally developed for machine vision applications, is proposed here as a postprocessing step to obtain two-dimensional velocity fields from spin-tagging MRI images. Results with artificially generated grids demonstrate the robustness of the optical flow algorithm to noise and indicate that a 7%–10% average error can be expected from the optical flow calculations alone, independent of MRI image artifacts. Experiments on spin-tagging MRI images for a Re=230 Poiseuille flow gave an average error of 6.41%, which was consistent with the measurement error of the generated (synthetic) images with the same level of random noise superimposed. © 2001 Biomedical Engineering Society.
PAC01: 8761-c, 8719Uv, 8757Ce
Similar content being viewed by others
REFERENCES
Bakker, C. J. G., R. Bhagwandien, M. A. Moerland, and M. Fuderer. Susceptibility artifacts in 2DFT spin-echo and gradient-echo imaging: The cylinder model revisited. Magn. Reson. Imaging 11:539–548, 1993.
Barron, J. L., D. J. Fleet, and S. S. Beauchemin. Performance of optical flow techniques. Int. J. Comput. Vis. 12:43–77, 1994.
Bittoun, J., E. Bourroul, O. Jolivet, I. Idy-Peretti, E. Mousseaux, A. Tardivon, and P. Peronneau. High-precision MR velocity mapping by 3D-Fourier phase encoding with a small number of encoding steps. Magn. Reson. Med. 29:674–680, 1993.
Black, M. J., and P. Anandan. The robust estimation of mul-iple motions: Parametric and piece-wise smooth flow fields. Comput. Visi. Image Understanding 63:75–104, 1996.
Browne, P., A. Ramuzat, R. Saxena, and A. P. Yoganathan. Experimental investigation of the steady flow downstream of the St. Jude bileaflet heart valve: A comparison between LDV and PIV techniques. Ann. Biomed. Eng. 28:39–47, 2000.
Callaghan, P. T. Principles of NMR Microscopy. Oxford: Clarendon, 1991.
Caprihan, A., and E. Fukushima. Flow measurements by NMR. Phys. Rep. 198:196–235, 1990.
Duerk, J. L., O. P. Simonetti, G. C. Hurst, and D. A. Finelli. Experimental conformation of phase encoding of instantaneous derivatives of position. Magn. Reson. Med. 32:77–87, 1994.
Feinauer, A., S. A. Altobelli, and E. Fukushima. NMR measurements of flow profiles in a coarse bed of packed spheres. Magn. Reson. Imaging 10:479–487, 1997.
Frayne, R., and B. K. Rutt. Understanding acceleration-induced displacement artifacts in phase-contrast MR velocity measurements. J. Magn. Resonan. Imaging 5:207–217, 1995.
Hill, R. B., and J. C. Klewicki. Data reduction methods for flow tagging velocity measurements. Exp. Fluids 20:863–871, 1996.
Hitt, D., M. L. Lowe, and R. Newcomer. Application of optical flow techniques to flow velocimetry. Phys. Fluids 7:6–8, 1995.
Jou, L.-D., R. van Tyen, S. A. Berger, and D. Saloner. Calculation of the magnetization distribution for fluid flow in curved vessels. Magn. Reson. Med. 35:577–584, 1996.
Klein, S. S., T. P. Graham, and C. H. Lorenz. Noninvasive delineation of normal right ventricular contractile motion with magnetic resonance imaging myocardial tagging. Ann. Biomed. Eng. 26:756–763, 1998.
Kutsovsky, Y. E., L. E. Scriven, and H. T. Davis. NMR imaging of velocity profiles and velocity distributions in bead packs. Phys. Fluids 8:863–871, 1996.
Kutter, E. MRI velocimetry in filtration flows: Time-of-flight vs. phase encoding methods, MS thesis, University of Illinois at Urbana-Champaign, 1998.
McVeigh, E. R. MRI of myocardial function: motion tracking techniques. Magn. Reson. Imaging 14:137–150, 1996.
Morris, P. G. NMR Imaging in Medicine and Biology. Oxford: Clarendon, 1986.
Moser, K. W., J. G. Georgiadis, and R. O. Buckius. On the accuracy of EPI-based phase contrast velocimetry. Magn. Reson. Imaging 18:1115–1123, 2000.
Moser, K. W., E. C. Kutter, J. G. Georgiadis, and R. O. Buckius. Velocity measurements of flow through a step stenosis using magnetic resonance imaging. Exp. Fluids 29:438–447, 2000.
Moser, K. W., L. G. Raguin, A. Harris, H. D. Morris, J. G. Georgiadis, M. Shannon, and M. Philpott. Visualization of Taylor-Couette and spiral Poiseuille flows using a snapshot FLASH spatial tagging sequence. Magn. Reson. Imaging 18:199–207, 2000.
Mosher, T. J., and M. B. Smith. A DANTE tagging sequence for the evaluation of translational sample motion. Magn. Reson. Med. 15:334–339, 1990.
Oshinski, M. S., D. N. Ku, D. E. Bohning, and R. I. Pettigrew. Effects of acceleration on the accuracy of MR phase velocity measurements. J. Magn. Reson. Imaging 2:665–670, 1992.
Pelc, N. J., M. A. Bernstein, A. Shimakawa, and G. H. Glover. Encoding strategies for three-direction phase-contrast MR imaging of flow. J. Magn. Reson. Imaging 1:405–413, 1991.
Shattuck, M., J. G. Behringer, G. A. Johnson, and J. G. Georgiadis. Convection and flow in porous media, I: Visual-ization by magnetic resonance imaging. J. Fluid Mech. 332:215–245, 1997.
Stuber, M., S. E. Fischer, M. B. Scheidegger, and P. Boesiger. Towards high resolution myocardial tagging. Magn. Reson. Med. 41:639–643, 1999.
Su, M.-Y., and O. Nalcioglu. Susceptibility effects in porous media in the presence of flow. J. Magn. Reson. Imaging 3:794–799, 1993.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Moser, K.W., Georgiadis, J.G. & Buckius, R.O. On the Use of Optical Flow Methods with Spin-Tagging Magnetic Resonance Imaging. Annals of Biomedical Engineering 29, 9–17 (2001). https://doi.org/10.1114/1.1332082
Issue Date:
DOI: https://doi.org/10.1114/1.1332082