The Edwards MIRA heart valve prosthesis: a 2-year study

J Card Surg. 2004 May-Jun;19(3):226-31. doi: 10.1111/j.0886-0440.2004.04060.x.

Abstract

Background: The Edwards MIRA mechanical heart valve is designed to optimize hemodynamics, reduce thrombogenicity, and avoid mechanical failure with a special hinge mechanism. The purpose of the study was to investigate the clinical performance and postoperative hemodynamic results of the first European patients receiving Edwards MIRA mechanical heart valves.

Methods: From March 1998 to March 1999 a total of 54 Edwards MIRA valves model numbers 3600 (aortic, n = 44) and 9600 (mitral, n = 10) were implanted in 52 (36 male, 16 female; mean age 61 +/- 10.1 years) consecutive patients undergoing mechanical valve replacement in a prospective study. Follow-up of the patients including physical examination, ECG, blood tests, and Doppler were performed prior to discharge, at 6 months, at 1 year, and at least 2 years postoperatively.

Results: Through October 2001 a total of 172 follow-up examinations were completed (51 patients at discharge, 46 patients at 6 months, 43 patients at 12 months, 32 patients at 2 years or beyond). All patients were in NYHA class I and II at the 6-month and 2+-year follow-up. All the patients stated an improved quality of life. Hospital mortality was 1.9%. There were no complications related to anticoagulation. Mean international normalized ratio at 6 months was 3.2 (range 1.9 to 4.3); lactate dehydrogenase was slightly increased with 264 +/- 103 U/L on average (normal value 80 to 240 U/L). No signs of valvular dysfunction or paravalvular leakage were observed. Mean pressure gradients were related to valve diameter: after mitral valve replacement (size 27, 29, 31 mm: 4.8, 3.2, 2.1 mmHg); after aortic valve replacement (size 19, 21, 23, 25 mm: 12.1, 13.1, 9.3, 8.2 mmHg).

Conclusions: These preliminary data suggest good hemodynamic function and a low rate of valve-related complications of the Edwards-MIRA mechanical prosthesis.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Aortic Valve / diagnostic imaging
  • Aortic Valve / pathology
  • Aortic Valve / surgery
  • Echocardiography
  • Female
  • Follow-Up Studies
  • Germany / epidemiology
  • Heart Valve Diseases / diagnostic imaging
  • Heart Valve Diseases / epidemiology
  • Heart Valve Diseases / surgery
  • Heart Valve Prosthesis Implantation
  • Heart Valve Prosthesis*
  • Humans
  • Male
  • Middle Aged
  • Mitral Valve / diagnostic imaging
  • Mitral Valve / pathology
  • Mitral Valve / surgery
  • Morbidity
  • Postoperative Complications / diagnostic imaging
  • Postoperative Complications / etiology
  • Postoperative Complications / mortality
  • Prosthesis Design*
  • Survival Analysis
  • Time Factors
  • Treatment Outcome