Doppler echocardiography of 119 normal-functioning St Jude Medical mitral valve prostheses: a comprehensive assessment including time-velocity integral ratio and prosthesis performance index

J Am Soc Echocardiogr. 2005 Mar;18(3):252-6. doi: 10.1016/j.echo.2004.11.011.

Abstract

The purpose of this study was to provide, in a large number of patients, comprehensive Doppler echocardiographic assessment of normal St Jude Medical mitral valve prosthesis function using Doppler-derived hemodynamic variables, including the mitral valve prosthesis-to-left ventricular outflow tract time-velocity integral ratio and prosthesis performance index. The pressure half-time was less than 130 milliseconds in all patients, and all but one patient had either a peak early mitral diastolic velocity of 2 m/s or less or a mitral valve prosthesis-to-left ventricular outflow tract time-velocity integral ratio of less than 2.2. There was a significant (P < .001) negative correlation between the prosthesis performance index and prosthesis size. This negative correlation suggests that there is more efficient use of the in vitro geometric orifice area with smaller prostheses.