Durability of the Hancock MO bioprosthesis compared with standard aortic valve bioprostheses

Ann Thorac Surg. 1995 Aug;60(2 Suppl):S221-8. doi: 10.1016/0003-4975(95)00253-h.

Abstract

To compare the durability of the Hancock modified orifice (Hancock MO, model 250 [H-MO]) valve with two other commonly used standard aortic valve bioprostheses, a cohort of 1,602 patients undergoing aortic valve replacement using porcine valves between 1971 and 1990 (excluding simultaneous mitral valve replacement) was analyzed retrospectively using Cox model multivariate techniques. Five hundred sixty-one patients received a composite H-MO valve, 652 received a standard Hancock model 242 (H) valve, and 389 received a Carpentier-Edwards model 2625 (C-E) valve. Mean age was 60 +/- 15 years (+/- 1 standard deviation) (71% male). Follow-up (10,247 patient-years) extended to 15 years and was 97% complete. The main focus of this study was bioprosthetic durability, using The American Association for Thoracic Surgery/The Society of Thoracic Surgeons guidelines to define structural valve deterioration (SVD). Multivariate analysis revealed that (younger) age (p < 10(-5), liver disease (p = 0.02), and 1981 to 1985 operative period (p = 0.012) were the only significant, independent predictors of SVD. In concordance with previous reports, the SVD freedom estimate was greater than 90% at 15 years for patients older than 70 years of age. Hepatic dysfunction had an adverse effect on SVD (estimated freedom from event at 10 years was 34 +/- 17% [standard error of mean] versus 78 +/- 2% for those without liver disease), but this affected only 3% of patients. Interestingly, one operative period (1981 to 1985) was associated with a slightly higher risk of SVD compared to the three other 5-year time windows. Valve type did not emerge as a significant risk factor for SVD.(ABSTRACT TRUNCATED AT 250 WORDS)

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Actuarial Analysis
  • Aortic Valve / surgery*
  • Bioprosthesis*
  • Cohort Studies
  • Female
  • Follow-Up Studies
  • Heart Valve Prosthesis* / adverse effects
  • Heart Valve Prosthesis* / mortality
  • Humans
  • Male
  • Middle Aged
  • Postoperative Complications
  • Prosthesis Failure
  • Reoperation
  • Retrospective Studies
  • Survival Rate